Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 15: 312-322, 1969;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mahoney, J. P.
Right arrow Articles by Small, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mahoney, J. P.
Right arrow Articles by Small, W.

Clinical Chemistry, Vol 15, 312-322, Copyright © 1969 by the American Association for Clinical Chemistry

Studies on Manganese

I. Determination in Serum by Atomic Absorption Spectrophotometry

John P. Mahoney 1, Kathleen Sargent 1, Margaret Greland 1, and Walter Small 1

1 Blood Research Laboratory, Department of Hematology, Carney Hospital, Boston, Mass. 02124.

Atomic absorption spectrophotometry offers a method applicable to the determination of serum manganese in clinical research. The signal-to-noise ratio is great enough to yield precise data even with 10X scale expansion. The method has the required sensitivity for the measurement of the concentrations encountered in human serum, even though the signals produced are very small. It requires a few manipulative steps, and the possibility of contamination is kept to a minimum. Interference from other materials usually present in the matrix does not present a problem.

The mean manganese concentration in the serum of normal fasting human subjects was 2.4 ± 0.7 µg/100 ml without any variation between the sexes. The concentration in the same individual does not vary over long periods of time. Initial observations do not indicate diurnal variation.

Large quantities of manganese administered to two normal individuals for 3 and 8 weeks, respectively, did not produce a detectable change in the serum manganese concentrations.

Studies are underway to adapt the method to the determination of manganese in blood cells and other tissues.

Submitted on April 3, 1968
Accepted on August 14, 1968







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1969 by the American Association for Clinical Chemistry.