Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 17: 103-107, 1971;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yee, H. Y.
Right arrow Articles by Bowles, F. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yee, H. Y.
Right arrow Articles by Bowles, F. R.

Clinical Chemistry, Vol 17, 103-107, Copyright © 1971 by the American Association for Clinical Chemistry

Modified Manual or Automated o-Toluidine System for Determining Glucose in Serum, with an Improved Aqueous Reagent

Hugh Y. Yee 1, Edward S. Jenest 1, and Fred R. Bowles 1

1 Department of Pathology, Hutzel Hospital, 432 E. Hancock St., Detroit, Mich. 48201.

A manual and automated method has been developed in which a new "aqueous" o-toluidine reagent is used that is compatible with standard tubing throughout the system. Citric acid, boric acid, and thiourea are incorporated, to increase absorbance readings. The sampling rate is 50/h, and the sample volume is 0.12 ml. Measurements may be made at either 630 or 385 nm.


Key Words: citric acid, boric acid, and thiourea additives

Submitted on August 17, 1970
Accepted on November 6, 1970