Clinical Chemistry, Vol 17, 229-230, Copyright © 1971 by the American Association for Clinical Chemistry
1 Department of Clinical Pathology and the Clinical
Research Center, University of Alabama Medical School and
University Hospital, Birmingham, Ala. 35233.
Changes are described in AutoAnalyzer Methodology N-24a, for the automated determination of serum total cholesterol, that increase the reaction temperature of the mixture of sample and acid color reagent. The modified procedure eliminates the inaccuracies that attended the original methodology and allows the cholesterol ester moiety in the serum total cholesterol to be quantitated.
Submitted on November 27, 1970