Clinical Chemistry, Vol 17, 610-613, Copyright © 1971 by the American Association for Clinical Chemistry
1 Department of Biochemistry, Memorial Hospital for
Cancer and Allied Diseases, New York, N. Y. 10021.
There is a well-recognized need for a reliable, convenient, and simple method for determining whole blood pH, PaCOaCO2, and PaOaO2. We have designed a specimen collection set that facilitates collection of the specimen and permits convenient and reliable analysis by the laboratory. Specimens are collected in newly designed "Vacutainer" tubes containing heparin and nitrogen gas. Without removing the Vacutainer stopper, the specimen may be anaerobically introduced into the analytical equipment through a unique aspirating needle. Atmospheric contamination of the specimen is eliminated.
Submitted on March 29, 1971
Accepted on May 5, 1971