Clinical Chemistry, Vol 17, 921-925, Copyright © 1971 by the American Association for Clinical Chemistry
1 Medical Laboratory Associates, 1025 S. 18th St., Birmingham, Ala. 35205.
A direct method is described, in which a single reagent is used without deproteinization, for measuring blood urea nitrogen (BUN) in serum. The reagent is a mixture of diacetylmonoxime, antipyrine (phenazone), and arsenic pentoxide in acid solution. Lipemia, bilirubin to 25 mg/100 ml, hemoglobin to lg/100 ml, and protein to 20g/100 ml do not affect the results. Beers law is followed at 460 nm for BUN concentrations as great as 100 mg/100 ml. The direct method compares favorably with a manual urease Berthelot color method and with the diacetylmonoxime procedure as automated for the Technicon SMA 6/60. The method (CV, 6.7%) can be completed in less than 20 mm. The reagent is stable for at least two months.
Submitted on May 31, 1971
Accepted on June 10, 1971
The following articles in journals at HighWire Press have cited this article:
![]() |
J. L. Ensunsa, J. D. Symons, L. Lanoue, H. R. Schrader, and C. L. Keen Reducing Arginase Activity via Dietary Manganese Deficiency Enhances Endothelium-Dependent Vasorelaxation of Rat Aorta Experimental Biology and Medicine, December 1, 2004; 229(11): 1143 - 1153. [Abstract] [Full Text] [PDF] |
||||