Clinical Chemistry, Vol 18, 630-642, Copyright © 1972 by the American Association for Clinical Chemistry
1 Biochemistry Department, Isaac Albert Research
Institute of the Kingsbrook Jewish Medical Center, Brooklyn,
N.Y. 11203.
Methodological modifications have been made in our "SMA 12/60 AutoAnalyzer" system, to improve the reproducibility and accuracy of six commonly performed procedures. These include glucose with the glucose oxidase method, albumin with the bromcresol green method, and alkaline phosphatase with the sodium thymolphthalein monophosphate method. We have also modified the composition of the reagents used in the determination of protein, inorganic phosphate, and aspartate aminotransferase. A general approach that is useful for adapting a new method to a multichannel AutoAnalyzer system is discussed.
Accepted on May 1, 1972