Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 18: 937-942, 1972;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Pelletier, O.
Right arrow Articles by Madère, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pelletier, O.
Right arrow Articles by Madère, R.

Clinical Chemistry, Vol 18, 937-942, Copyright © 1972 by the American Association for Clinical Chemistry

New Automated Method for Measuring Thiamine (Vitamin B1) in Urine

Omer Pelletier 1 and René Madère 1

1 Nutrition Research Division, Health Protection Branch, Tunney’s Pasture, Ottawa, Ontario K1A 0L2, Canada.

A new automated method was developed to eliminate the time-consuming steps of the manual procedure for the measurement of thiamine (vitamin B1) in urine. A Technicon Pump II simultaneously pumped 23 samples through chromatographic columns of "Decalso," washed out the interfering substances, and eluted thiamine. The automated procedure for determining thiamine with a Technicon AutoAnalyzer system involved the oxidation of thiamine to thiochrome, extraction into isobutanol and fluorescence measurements. Results obtained by the automated method correlated well with those obtained by the manual procedure. Recoveries of added thiamine were satisfactory and the results reproducible. This method was used successfully in the analysis of more than 10,000 urine samples for the Canadian Nutrition Survey.


Key Words: AutoAnalyzer • nutrition status assessment • "Decalso" columns • fluorometry

Submitted on March 31, 1972
Accepted on June 12, 1972