p 2: A statement is made under "Principle" that is not consistent with the "Procedure:" "At zero time and at 60 min, aliquots are taken, deproteinized with trichloroacetic acid, and inorganic phosphorus determined in aliquots of the filtrate." There is no step in the "Procedure" that indicates removal of aliquots at zero time.
p 5: "Normal Values and Precision." The last sentence indicates a higher value for inorganic phosphate in the tube inhibited with nickel than in the tube without the inhibitor. The reverse is true.