Clinical Chemistry, Vol 20, 1454-1457, Copyright © 1974 by the American Association for Clinical Chemistry
1 B. C. Bio-Medical Laboratories Ltd., 7845 Edmonds St., Burnaby, B. C., Canada.
We describe a completely automated enzymatic method for estimating serum triglycerides with a centrifugal analyzer (ENI-GEMSAEC). Accuracy and precision (CV, 2 to 4.8%) and the upper limit of linearity (400 mg/100 ml) are monitored by use of newly developed, clear, aqueous triolein standards. A modified computer program (FOCAL 8) is used for data manipulation and correcting absorbance to zero reaction time. Results cornpare well with those obtained by an automated fluorometric procedure and a manual enzymatic procedure.
Submitted on July 1, 1974