Clinical Chemistry, Vol 21, 672-675, Copyright © 1975 by the American Association for Clinical Chemistry
1 Pharmacology Research Laboratory (151), Veterans Administration Center, Wood (Milwaukee), Wis. 53193; and the Department
of Pharmacology, Medical College of Wisconsin, Milwaukee, Wis.
53233.
A radioimmunoassay (I) for barbiturates was compared with thin-layer chromatographic (II) and gasliquid chromatographic (III) methods for barbiturate detection in human urine. Timed urine samples were obtained from volunteers who had ingested 100 mg of a barbiturate. I detected barbiturate in all urines tested up to 76 h after the dose, and III in all up to 52 h and in 90% up to 76 h. II detected barbiturates in 90% of all urine samples for only 30 h, after which its reliability declined. Glutethimide interfered with radioimmunoassay of barbiturate, producing false positives. I is sensitive, reliable, and fast, and lends itself to screening large numbers of urine samples for barbiturates. For routine urine surveillance, however, we found I to be less useful than II, which is still the method of choice. I has, however, proved to be an excellent method for confirming resuits of II.
Submitted on November 18, 1974
Accepted on February 3, 1975