Clinical Chemistry, Vol 21, 850-852, Copyright © 1975 by the American Association for Clinical Chemistry
1 Department of Pathology, Upstate Medical Center State University of New York, 750 E. Adams St., Syracuse, N. Y. 13210.
We evaluated a revised serum calcium method (o-cresolphthalein complexone) used with the Du Pont aca. Our results correlated well with those obtained by atomic absorption spectrometry. Within-run and day-to-day coefficients of variation were 0.5 and 1.5%, respectively. Color produced and calcium concentration are linearly related and the relation is not affected by hemolysis, lipemia, or icterus.
Submitted on January 24, 1975
Accepted on March 24, 1975