Clinical Chemistry, Vol 22, 1589-1592, Copyright © 1976 by American Association for Clinical Chemistry
DW Seccombe, P Dodek, J Frohlich, P Hahn, JP Skala and DJ Campbell
Because of renewed interest in a possible connection between carnitine, lipid disorders, and myopathy, an automated method of analysis is desirable. Deproteinization of serum by use of membrane filter cones and automated assay with a bichromatic analyzer (the ABA-100) substantially increases efficiency without sacrificing the specificity and accuracy of the original manual enzymatic method. The described procedure allows for analysis of 80 speciments a day and is thus suitable for screening of selected populations. Normal values found in blood sera of adults were in the range of 25.0-73.8 mu mol/liter and the method has sufficient sensitivity to accurately measure concentrations as small as 10 mu mol/liter.