J Vasiliades
Spectrophotometric, kinetic, and nuclear magnetic resonance studies indicate that alkaline sodium picrate and creatinine react to form a 1/1 aduct between picric and creatinine, with a stability constant of log K= 4.26. Kinetic studies indicate that the forward reaction is first order with respect to picric acid, hydroxide, and creatinine concentration. The reverse reaction, the dissociation of the 1/1 complex, shows a complex dependence on hydroxide concentration. The expression for the observed pseudo-first-order rate constant in the presence of excess picric acid is: Kobsd = K1K0[P][OH] +[K2[OH]x. A value of K1K0 = 5.0 (mol/liter)-2s-1 is obtained. For accurate analytical results with this reaction, hydroxide concentration must be maintained at a constant value for both samples and standards.
The following articles in journals at HighWire Press have cited this article:
![]() |
J. M Ellis, H. K. Tan, R. E Gilbert, D. P R Muller, W. Henley, R. Moy, R. Pumphrey, C. Ani, S. Davies, V. Edwards, et al. Supplementation with antioxidants and folinic acid for children with Down's syndrome: randomised controlled trial BMJ, March 15, 2008; 336(7644): 594 - 597. [Abstract] [Full Text] [PDF] |
||||