Clinical Chemistry, Vol 22, 350-358, Copyright © 1976 by American Association for Clinical Chemistry
D Rodbard, RH Lenox, HL Wray and D Ramseth
We have developed practical methods for evaluating the magnitude of the random errors in radioimmunoassay dose--response variables, and the relationship between this error and position on the dose--response curve. This is important: to obtain appropriate weights for each point on the dose--response curve when utilizing least-squares curve-fitting methods; to evaluate whether the standards and the unknowns are subject to error of the same magnitude; for quality-control purposes; and to study the sources of errors in radioimmunoassay. Both standards and unknowns in radioimmunoassays for cAMP and cGMP were analyzed in triplicate. The same mean (Y), sample standard deviation, sy, and variance (2-y) of the response variable were calculated for each dose level. The relationship between s 2-y and y was calculated utilizing several models. Results for standards and unknowns from several assays were pooled, and a curve smoothing procedure was used to minimize random sampling errors. This pooling increased the reliability of the analysis, and confirmed the presence of the theoretically predicted nonuniformity of variance. Thus, the calculation of results from these radioimmunoassays should utilize a weighted least-squares curve-fitting program. These analyses have been computerized, and can be used as a "pre-processor" for programs for routine analysis of results of radioimmunoassay.
The following articles in journals at HighWire Press have cited this article:
![]() |
D. Curran-Everett Estimation of dose-response curves and identification of peaks in hormone pulsations: classic marriages of statistics to science Am J Physiol Endocrinol Metab, September 1, 2005; 289(3): E363 - E365. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Wang, D. H. Catlin, L. M. Demers, B. Starcevic, and R. S. Swerdloff Measurement of Total Serum Testosterone in Adult Men: Comparison of Current Laboratory Methods Versus Liquid Chromatography-Tandem Mass Spectrometry J. Clin. Endocrinol. Metab., February 1, 2004; 89(2): 534 - 543. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Straume, M. L. Johnson, and J. D. Veldhuis Statistically accurate estimation of hormone concentrations and associated uncertainties: methodology, validation, and applications Clin. Chem., January 1, 1998; 44(1): 116 - 123. [Abstract] [Full Text] [PDF] |
||||