Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 23: 2275-2278, 1977;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Maitra, S. K.
Right arrow Articles by Guze, L. B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Maitra, S. K.
Right arrow Articles by Guze, L. B.

Clinical Chemistry, Vol 23, 2275-2278, Copyright © 1977 by American Association for Clinical Chemistry

Serum gentamicin assay by high-performance liquid chromatography

SK Maitra, TT Yoshikawa, JL Hansen, I Nilsson-Ehle, WJ Palin, MC Schotz and LB Guze

We describe a high-performance liquid chromatographic method for the quantitative determination of gentamicin in serum. The antibiotic was separated from serum by passage through a silicic acid column, derivatized with o-phthalaldehyde, and eluted with ethanol. The derivatized gentamicin was then separated into all three of its major components by reversed-phase chromatography and quantified by fluorometry. Concentrations in serum as low as 0.5 mg of gentamicin per liter could be accurately determined. A standard curve showed a linear response for serum containing gentamicin at concentrations ranging from 0 to 20 mg/liter. Tobramycin, amikacin, ampicillin, penicillin G, methicillin, carbenicillin, chloramphenicol, clindamycin, and cephalothin did not interfere with the gentamicin assay. Comparison with an accepted microbiological assay yielded a correlation coefficient of 0.99. This chemical assay is rapid (less than 30 min), sensitive, accurate, specific, and appears to be applicable to other aminoglycosides.


The following articles in journals at HighWire Press have cited this article:


Home page
Proc. Natl. Acad. Sci. USAHome page
C. L. Bowe, L. Mokhtarzadeh, P. Venkatesan, S. Babu, H. R. Axelrod, M. J. Sofia, R. Kakarla, T. Y. Chan, J. S. Kim, H. J. Lee, et al.
Design of compounds that increase the absorption of polar molecules
PNAS, October 28, 1997; 94(22): 12218 - 12223.