Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 24: 1813-1817, 1978;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Sundaram, P. V.
Right arrow Articles by Hinsch, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sundaram, P. V.
Right arrow Articles by Hinsch, W.

Clinical Chemistry, Vol 24, 1813-1817, Copyright © 1978 by American Association for Clinical Chemistry

Immobilized-enzyme nylon-tube reactor for routine determination of uric acid in serum

PV Sundaram, MP Igloi, R Wassermann and W Hinsch

We report here the preparation of an immobilized-enzyme nylon-tube reactor containing uricase (EC 1.7.3.3) and the assembly of a flow- through system (Technicon AutoAnalyzer II) for the routine determination of uric acid in serum. Results of these uric acid analyses by use of immobilized uricase, in conjunction with peroxidase (EC 1.11.1.7) and aminophenazone-dichlorophenol in solution, are compared with those obtained with the same enzyme in solution by use of the uricase-PAP (peroxidase, 4-aminophenazone, dichlorophenol) method. Clinical trials carried out routinely with the uricase reactor give reliable and reproducible results with high precision at an appreciably lower cost. The reactors are stable to continued or intermittent use for at least three months or for 4000 tests.


The following articles in journals at HighWire Press have cited this article:


Home page
ScienceHome page
R. Elin
Instrumentation in clinical chemistry
Science, October 17, 1980; 210(4467): 286 - 289.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1978 by the American Association for Clinical Chemistry.