Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 24: 374-376, 1978;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Fudge, A. N.
Right arrow Articles by Fraser, C. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Fudge, A. N.
Right arrow Articles by Fraser, C. G.

Clinical Chemistry, Vol 24, 374-376, Copyright © 1978 by American Association for Clinical Chemistry

Evaluation of a radiometric method for determination of total and unsaturated iron-binding capacities and serum iron

AN Fudge and CG Fraser

We describe an evaluation of the "RES-OMAT FE-59" radiometric kit (Mallinckrodt Australia Pty. Ltd.) for determination of total iron- binding capacity and unsaturated iron-binding capacity; serum iron may be calculated from total iron-binding capacity and unsaturated iron- binding capacity. Values obtained on patients' sera were compared with those obtained by continuous-flow analysis; correlation coefficients were 0.92, 0.93, and 0.88 for total iron-binding capacity, unsaturated iron-binding capacity, and iron, respectively. Precision between-run and within-run was almost identical and compares very favorably with that reported by Dixon [Ann. Clin. Biochem. 10, 127 (1973)] and that achieved in inter-laboratory surveys. Analytical recovery averaged only 83.5%, but results of analyses of sera with assigned or consensus values showed good accuracy. Satisfactory performance cannot be achieved with lyophilzed bovine quality-control materials. The kit has significant technical advantages over colorimetric methods and good short- and long-term analytical performance.





HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1978 by the American Association for Clinical Chemistry.