Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 24: 944-946, 1978;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Castro, A.
Right arrow Articles by Malkus, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Castro, A.
Right arrow Articles by Malkus, H.

Clinical Chemistry, Vol 24, 944-946, Copyright © 1978 by American Association for Clinical Chemistry

A totally automated system for enzyme immunoassay of theophylline in serum

A Castro, J Ibanez, W Voight, T Noto and H Malkus

We describe a procedure for enzyme immunoassay of theophylline (1,3- dimethylxanthine) in which all phases of the assay are totally automated in a Kinetic Analyzer (KA-150). This system permits assay of 75 10-mul samples per hour, with results available at 30-s intervals after initial sample preparation and preincubation. We compared results for 138 clinical samples by an ultraviolet method (x) and the present method (y). The slope of the comparison curve was 0.902, the y- intercept 0.402, and the correlation coefficient 0.984. The coefficient of variation for samples run in duplicate on the same day was 4.9%; it was 8.1% for samples run on different days. Specificity, sensitivity, simplicity, speed, and small reagent requirement all make this an attractive alternative to chromatographic procedures.