Clinical Chemistry 24: 1366-1372, 1978;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Fogt, E. J.
Right arrow Articles by Clemens, A. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Fogt, E. J.
Right arrow Articles by Clemens, A. H.

Clinical Chemistry, Vol 24, 1366-1372, Copyright © 1978 by American Association for Clinical Chemistry

Development and evaluation of a glucose analyzer for a glucose controlled insulin infusion system ((Biostator)

EJ Fogt, LM Dodd, EM Jenning and AH Clemens

The Glucose-Controlled Insulin Infusion System (Biostator) is a modular, computerized, feedback control system for dynamic control of blood glucose concentrations in diabetics. This on-line glucose analyzer for use with whole blood utilizes a novel enzyme (glucose oxidase)-membrane configuration and an electrochemical cell to measure the H202 generated. The analyzer exhibits both short- and long-range stability, and instrument response and analyte concentration are linearly related over the full range of clinical interest. The response is fast, accurate, and precise, and permits determination of blood glucose within 2 min from the moment the blood leaves the patient. Correlation studies were completed to show the agreement between the Biostator Glucose Analyzer and the FDA's recommended hexokinase/glucose- 6-phosphate dehydrogenase procedure on whole blood (e.g., average per cent recovered for 11 concentrations between 250 and 900 mg/liter was: hexokinase, 95.6%, Biostator Analyzer, 95.9%; bias and SDd, respectively, at low, normal, and high glucose values were: 12 and 41 mg/liter at the 500 mg/liter level; 4 and 52 mg/liter at the 1000 mg/liter level, and 4 and 128 mg/liter at the 4000 mg/liter level). No appreciable interference is observed with above-normal concentrations of bilirubin, uric acid, creatinine, sodium salicylate, or dextran. Platelet adhesion, which tends to decrease the useful life of the membrane, has been significantly decreased.


The following articles in journals at HighWire Press have cited this article:


Home page
DiabetesHome page
G. M. Steil, K. Rebrin, C. Darwin, F. Hariri, and M. F. Saad
Feasibility of Automating Insulin Delivery for the Treatment of Type 1 Diabetes
Diabetes, December 1, 2006; 55(12): 3344 - 3350.
[Abstract] [Full Text] [PDF]


Home page
Hum ReprodHome page
D. Cibula, M. Fanta, M. Hill, G. Sindelka, J. Skrha, and J. Zivny
Insulin sensitivity in non-obese women with polycystic ovary syndrome during treatment with oral contraceptives containing low-androgenic progestin
Hum. Reprod., January 1, 2002; 17(1): 76 - 82.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1978 by the American Association for Clinical Chemistry.