Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 25: 117-121, 1979;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Eisenwiener, H. G.
Right arrow Articles by Keller, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Eisenwiener, H. G.
Right arrow Articles by Keller, M.

Clinical Chemistry, Vol 25, 117-121, Copyright © 1979 by American Association for Clinical Chemistry

Absorbance measurement in cuvettes lying longitudinal to the light beam

HG Eisenwiener and M Keller

Measurement in cuvettes with their longitudinal dimension parallel to the light beam, while under the influence of a centrifugal force, has many advantages that are not immediately apparent. This survey summarizes the advantages in comparison to cuvettes standing with their long axis perpendicular to the light beam: in the latter the pathlength is fixed by the geometry, but if the cuvette is lying longitudinal to the light beam the pathlength is given by the filling volume. Because of the dependence of the pathlength on the filling (reaction) volume when cuvettes are lying longitudinal to the light beam, the calculation formulas show that the calculation factors, both for enzyme and substrate determination, become independent of the reagent volume form. This is demonstrated by simple equations and confirmed in practice by use of the (Cobas) Bio centrifugal analyzer in assays for glucose and aspartate aminotransferase.


The following articles in journals at HighWire Press have cited this article:


Home page
ScienceHome page
R. Elin
Instrumentation in clinical chemistry
Science, October 17, 1980; 210(4467): 286 - 289.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1979 by the American Association for Clinical Chemistry.