Clinical Chemistry, Vol 25, 156-158, Copyright © 1979 by American Association for Clinical Chemistry
B Vinet and L Zizian
We describe a method for the 7-min determination of theophylline in less than 100 muL of serum. The procedure requires no centrifugation or solvent evaporation after extraction. Butylation is done on the gas- chromatographic column by injecting the serum extract followed by a butylating mixture which contains 1-iodobutane as the alkylating agent. The method is precise, accurate, and free of interference. Results correlate well with those by ultraviolet spectrophotometry.