Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 25: 322-324, 1979;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Shaw, E. J.
Right arrow Articles by Smith, D. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Shaw, E. J.
Right arrow Articles by Smith, D. S.

Clinical Chemistry, Vol 25, 322-324, Copyright © 1979 by American Association for Clinical Chemistry

Continuous-flow fluoroimmunoassay of serum gentamicin, with automatic sample blank correction

EJ Shaw, RA Watson and DS Smith

We describe the continuous-flow automation of a quenching fluoroimmunoassay for determination of the antibiotic gentamicin in serum. In the flow system, sample is mixed with fluorescein-labeled gentamicin, followed by antiserum, and after a short incubation the fluorescence is measured. The intrinsic blank signal of each sample is corrected for during its single passage through the system by pumping antiserum discontinuously, so as to divide the final stream into sections corresponding to conventional separate assay and blank mixtures. Analyses time for each sample is 6 min, and 20 samples can be run per hour. Sensitivity, precision, and accuracy are adequate for clinical purposes, and assays of patients' samples correlate satisfactorily with the results of manual quenching and polarization fluoroimmunoassays, and plate diffusion bioassay.