Clinical Chemistry, Vol 25, 665-668, Copyright © 1979 by American Association for Clinical Chemistry
D Riad-Fahmy, GF Read, SJ Gaskell, J Dyas and R Hindawi
A simple, direct radioimmunoassay for cortisol in human serum and plasma is described. An antiserum, raised in sheep to a cortisol-3-(O- carboxymethyl)oxime/bovine serum albumin conjugate, is coupled to microcellulose. No extraction is required because plasma samples and standards are incubated with the antiserum and an 125I radioligand in a low-pH buffer, which denatures cortisol-binding globulins. The assay satisfies accepted validation criteria. In addition, results from the radioimmunoassay compare well with those obtained by a gas chromatographic-mass spectrometric technique (r = 0.968; FRIA = 0.97 FGCMS + 2.0 nmol/L). The latter procedure features the very high intrinsic specificity obtained by selected ion monitoring at high mass- spectrometric resolution (M/deltaM = 8500) with a Varian MAT-731 instrument. The simplicity of the radioimmunoassay procedure, with use of reagents prepared "in house," makes this a very practical and economical assay for use in the medium or large endocrine laboratory.
The following articles in journals at HighWire Press have cited this article:
![]() |
G. D. Smith, Y. Ben-Shlomo, A. Beswick, J. Yarnell, S. Lightman, and P. Elwood Cortisol, Testosterone, and Coronary Heart Disease: Prospective Evidence From the Caerphilly Study Circulation, July 19, 2005; 112(3): 332 - 340. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Hack, S. W. Lockley, J. Arendt, and D. J. Skene The Effects of Low-Dose 0.5-mg Melatonin on the Free-Running Circadian Rhythms of Blind Subjects J Biol Rhythms, October 1, 2003; 18(5): 420 - 429. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |