Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 26: 314-317, 1980;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Alexander, R. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Alexander, R. L., Jr

Clinical Chemistry, Vol 26, 314-317, Copyright © 1980 by American Association for Clinical Chemistry

Comparison of radial immunodiffusion and laser nephelometry for quantitating some serum proteins

RL Alexander Jr

After quantitating immunoglobulins G, A, and M and complement C3c and C4 in serum by using a laser nephelometer coupled with a data processor, I compared these results with values obtained by an early- readout radial immunodiffusion method. Day-to-day precision was better for nephelometry than for radial immunodiffusion for all proteins analyzed. The average coefficient of variation was 6.0% for nephelometry and 9.9% for radial immunodiffusion. Comparison of these methods gave ranked correlation coefficients of 0.945, 0.981, 0.932, 0.803, and 0.792 for IgG, IgA, IgM, C3c, and C4, respectively. Nephelometry gave significantly higher values than radial immunodiffusion for IgG, IgA, IgM, and C3c, and significantly lower values for C4 (p less than 0.001). Part of this bias was found to be due to the equation programmed in the data processor for calculating the standard curves. Within 95% limits, nephelometry gave higher normal ranges than radial immunodiffusion for IgG, IgA, and IgM. Other possible factors that can produce this bias are discussed.


The following articles in journals at HighWire Press have cited this article:


Home page
ScienceHome page
R. Elin
Instrumentation in clinical chemistry
Science, October 17, 1980; 210(4467): 286 - 289.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1980 by the American Association for Clinical Chemistry.