Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 26: 763-765, 1980;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Baxter, R. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Baxter, R. C.

Clinical Chemistry, Vol 26, 763-765, Copyright © 1980 by American Association for Clinical Chemistry

Simplified approach to confidence limited in radioimmunassay

RC Baxter

A simple method of calculating confidence limits for radioimmunoassay data is presented. The method involves the use of the within-assay variation in dose estimate of three routine quality-control specimens, measured in repeated assays, to estimate the confidence limits for results on unknown samples. Results for control specimens are combined by calculating the unique quadratic curve fitting a graph of within- assay standard deviation vs mean value for each control. This method requires no special data accumulation or advanced computing equipment. For cortisol, lutropin, and thyroxine radioimmunoassays, confidence limits calculated in this way have been compared with those calculated from the variance of the response variable "B/B0" in repeated standard curves. Both methods agree well with actual limits observed when plasma pools containing a wide range of hormone concentrations are assayed repeatedly.


The following articles in journals at HighWire Press have cited this article:


Home page
Ann Clin BiochemHome page
W. A Sadler
Error models for immunoassays
Ann Clin Biochem, September 1, 2008; 45(5): 481 - 485.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1980 by the American Association for Clinical Chemistry.