Clinical Chemistry, Vol 27, 35-38, Copyright © 1981 by American Association for Clinical Chemistry
T Ono, K Kitaguchi, M Takehara, M Shiiba and K Hayami
We studied the effects on 25 analytes of duration of contact of serum with non-anticoagulated blood and of temperature. Serum was separated after blood was allowed to stand, for 0, 2, 4, 6, 8, 24, or 48 h at 4, 23, or 30 degrees C. Results obtained for bilirubin, albumin, zinc sulfate turbidity, thymol turbidity, cholinesterase (EC 3.1.1.8), alkaline phosphatase (EC 3.1.3.1), leucine aminopeptidase (EC 3.4.11.1), amylase (EC 3.2.1.2), total cholesterol, triglycerides, beta- lipoprotein, serum urea nitrogen, creatinine, uric acid, and gamma- glutamyltransferase (EC 2.3.2.2) were not influenced by storage at 4, 24, or 30 degrees C for as long as 48 h. Negligible differences were seen for potassium in sera in contact with cells as long as 24 h at 23 degrees C and for inorganic phosphorus after 48 h at 4 degrees C. However, at 4 degrees C we noted an increase at 8 h, a slight decrease at 30 degrees C. Statistically significant changes were seen for total protein and calcium after 48 h at 30 degrees C; for aspartate aminotransferase (EC 2.6.1.1), and alanine aminotransferase (EC 2.6.1.2), between 8 and 24 h at 23 degrees C and as soon as 6 h at 30 degrees C; for lactate dehydrogenase (EC 1.1.1.27) after 8 h at 30 degrees C and between 8 and 24 h at 23 degrees C; for glucose at 24, 4, or 2 h of storage at 4, 23, or 30 degrees C, respectively; for inorganic phosphorus after 48 h at 23 degrees C or 8 h at 30 degrees C; for potassium after 4 h at 4 degrees C or 24 h at 30 degrees C; and for sodium after 48 h at 4 degrees C or 6 h at 23 or 30 degrees C.
The following articles in journals at HighWire Press have cited this article:
![]() |
M. Tanner, N. Kent, B. Smith, S. Fletcher, and M. Lewer Stability of common biochemical analytes in serum gel tubes subjected to various storage temperatures and times pre-centrifugation Ann Clin Biochem, July 1, 2008; 45(4): 375 - 379. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.-A. von Mach, L. S. Weilemann, and P. von Landenberg Pseudocholinesterase Activity in Organophosphate Poisoning after Storage of Unseparated Blood Samples at Room Temperature for 3 Weeks Clin. Chem., October 1, 2005; 51(10): 1907 - 1908. [Full Text] [PDF] |
||||
![]() |
E. J. Giltay, J. M. Geleijnse, E. G. Schouten, M. B. Katan, and D. Kromhout High Stability of Markers of Cardiovascular Risk in Blood Samples Clin. Chem., April 1, 2003; 49(4): 652 - 655. [Full Text] [PDF] |
||||
![]() |
S. Clark, L. D Youngman, A. Palmer, S. Parish, R. Peto, and R. Collins Stability of plasma analytes after delayed separation of whole blood: implications for epidemiological studies Int. J. Epidemiol., February 1, 2003; 32(1): 125 - 130. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L. Boyanton Jr and K. E. Blick Stability Studies of Twenty-Four Analytes in Human Plasma and Serum Clin. Chem., December 1, 2002; 48(12): 2242 - 2247. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Dufour, J. A. Lott, F. S. Nolte, D. R. Gretch, R. S. Koff, and L. B. Seeff Diagnosis and Monitoring of Hepatic Injury. I. Performance Characteristics of Laboratory Tests Clin. Chem., December 1, 2000; 46(12): 2027 - 2049. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Zhang, R. K. Elswick, W. G. Miller, and J. L. Bailey Effect of serum-clot contact time on clinical chemistry laboratory results Clin. Chem., June 1, 1998; 44(6): 1325 - 1333. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |