Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 27: 1753-1757, 1981;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Flynn, S. D.
Right arrow Articles by Grauds, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Flynn, S. D.
Right arrow Articles by Grauds, S.

Clinical Chemistry, Vol 27, 1753-1757, Copyright © 1981 by American Association for Clinical Chemistry

Factors affecting enzyme-linked immunosorbent assay (ELISA) for insulin antibodies in serum

SD Flynn, DF Keren, B Torretti, RC Dieterle and S Grauds

We have evaluated the factors affecting an enzyme-linked immunosorbent assay (ELISA) for circulating insulin antibodies. Dilutions of patients' sera were incubated in polystyrene tubes coated with procine insulin. The second incubation was with alkaline phosphatase-conjugated anti-human Fab. Each of these reactions was complete after 4 h. Specificity of the reaction for insulin antibodies was demonstrated by removal of anti-insulin activity after preincubation with insulin, but not with glucagon at similar concentrations. Sensitivity of the ELISA system, assessed by performing the reaction with affinity-column purified insulin antibodies, was 10 microgram of specific antibody per liter. Using this system, we examined sera from 22 patients who had been determined by radioimmunoassay to have insulin antibodies, and sera from 23 normal individuals. The ELISA results correlated reasonably well with those of RIA (r = +0.84). Besides detecting insulin antibodies in diabetic patients who are being treated with insulin, we use this ELISA test as a screening procedure to be certain insulin antibodies are not present when we use indirect immunofluorescence methods of fixed pancreatic substrate to detect islet-cell antibodies.


The following articles in journals at HighWire Press have cited this article:


Home page
Endocr. Rev.Home page
S. E. Fineberg, T. T. Kawabata, D. Finco-Kent, R. J. Fountaine, G. L. Finch, and A. S. Krasner
Immunological Responses to Exogenous Insulin
Endocr. Rev., October 1, 2007; 28(6): 625 - 652.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1981 by the American Association for Clinical Chemistry.