Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 27: 1866-1871, 1981;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Subramanian, K. S.
Right arrow Articles by Meranger, J. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Subramanian, K. S.
Right arrow Articles by Meranger, J. C.

Clinical Chemistry, Vol 27, 1866-1871, Copyright © 1981 by American Association for Clinical Chemistry

A rapid electrothermal atomic absorption spectrophotometric method for cadmium and lead in human whole blood

KS Subramanian and JC Meranger

A rapid graphite-furnace atomic absorption spectrophotometric procedure is described for determining cadmium and lead in heparinized human whole blood. A known aliquot of the blood sample is diluted fivefold with an aqueous solution composed of 5 g each of diammonium hydrogen phosphate and Triton X-100 per liter, the solution is vigorously agitated, and a 10-microL aliquot is injected into a pyrocoated graphite tube under optimized instrumental conditions. Values for Cd and Pb in the sample are obtained by direct comparison to linear working curves prepared from aqueous standards of the metals in the diammonium hydrogen phosphate-Triton medium; there is no need to use the method of standard addition or matrix-matched calibration curves. Also, the method is free of matrix effects. At least 30 samples can be analyzed per hour. The rapidity, simplicity, and sensitivity of the method make it attractive as a screening technique for routine environmental surveillance involving large throughput of samples.


The following articles in journals at HighWire Press have cited this article:


Home page
Asia Pac J Public HealthHome page
J. Sirivarasai, S. Kaojarern, W. Wananukul, W. Deechakwan, and P. Srisomerarn
Non-occupational Lead and Cadmium Exposure and Blood Pressure in Thai Men
Asia Pac J Public Health, January 1, 2004; 16(2): 133 - 137.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1981 by the American Association for Clinical Chemistry.