Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 28: 2106-2109, 1982;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Robertson, E. A.
Right arrow Articles by Elin, R. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Robertson, E. A.
Right arrow Articles by Elin, R. J.

Clinical Chemistry, Vol 28, 2106-2109, Copyright © 1982 by American Association for Clinical Chemistry

Lactate dehydrogenase activity as measured by the SMAC is related to flow-cell energy

EA Robertson, E Wright, RA Chesler and RJ Elin

Determination of lactate dehydrogenase (LDH) activity in the SMAC (Technicon) is based on the change in NADH absorbance between two flow cells. We noted that results for patients' specimens and controls changed when the fiber optic terminations for the two LDH channel flow cells were adjusted or "peaked" at the colorimeter chopper assembly. The energy (intensity) of light reaching the flow cells was varied by adjusting the fiber optic terminations, and the absorbance readings for a series of solutions containing NADH and patients' specimens were recorded. For both flow cells, when the fiber optic terminations were adjusted to increase the zero absorbance light intensity from 20 lines to 60 lines, a significant (p less than 0.0001) proportional change was seen in the absorbance readings. Evidently the difference in absorbance between the two flow cells is related not only to the NADH concentrations but also to the difference in the light intensity at the two flow cells. Consequently, changes in the adjustment of the fiber optic terminations produce systematic changes in results for LDH in patients' sera. These systematic changes in LDH results may be minimized by maintaining equivalent settings of the fiber optic terminations for the two flow cells and by using the calibration material with an absorbance most similar to that of patients' specimens.