Clinical Chemistry, Vol 28, 2421-2422, Copyright © 1982 by American Association for Clinical Chemistry
R Balamut, D Doran, G Giridhar, D Mitchell and S Soule
Hematofluorometers manufactured by two different companies (Aviv Biomedical and Environmental Science Associates) performed satisfactorily for protoporphyrin determinations with proficiency test samples but showed about a 30% low bias with actual clinical blood specimens.
The following articles in journals at HighWire Press have cited this article:
![]() |