Clinical Chemistry, Vol 28, 502-504, Copyright © 1982 by American Association for Clinical Chemistry
TR Koch, PK Adolf, DH Zimmerman, EC Knoblock and VM Leitz
For these quantitative turbidimetric assays for immunoglobulins IgG, IgA, and IgM in serum, the reagents used are commercially available in kit form. The two-point assays are performed with a miniature centrifugal analyzer (GEMENI, Electro-Nucleonics, Inc.). The reading taken 5 s after mixing is used to correct for specimen turbidity. Run- to-run precision (CV) was typically 5% (range: 3.0-8.7%). Excellent correlation with radial immunodiffusion was observed for each test (r greater than 0.98 for each test). The dynamic range of each test extends above and below the reference-interval ranges.