Clinical Chemistry, Vol 28, 1192-1196, Copyright © 1982 by American Association for Clinical Chemistry
TG Brewster, DJ Waite and GA Hudson
We measured beta-lipoprotein in 232 cord sera by rate nephelometry and compared the results with those obtained by electroimmunoassay. The mean (mean +/- two standard deviations) was 544 (293-1009) mg/L by rate nephelometry and 503 (265-955) mg/L by electroimmunoassay. The between- assay correlation coefficient was 0.891. Between-run coefficients of variation were 3.5-4.4% for nephelometry and 11-14% for electroimmunoassay. Nephelometry was faster, more precise, and less labor intensive than electroimmunoassay. For these reasons, rate nephelometry appears to be a more suitable assay system for mass screening, should detection of familial hypercholesterolemia prove feasible and desirable.
The following articles in journals at HighWire Press have cited this article:
![]() |
E. M. Haney, L. H. Huffman, C. Bougatsos, M. Freeman, R. D. Steiner, and H. D. Nelson Screening and Treatment for Lipid Disorders in Children and Adolescents: Systematic Evidence Review for the US Preventive Services Task Force Pediatrics, July 1, 2007; 120(1): e189 - e214. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Vuorio, H. Turtola, and K. Kontula Neonatal Diagnosis of Familial Hypercholesterolemia in Newborns Born to a Parent With a Molecularly Defined Heterozygous Familial Hypercholesterolemia Arterioscler Thromb Vasc Biol, November 1, 1997; 17(11): 3332 - 3337. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |