Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 28: 1514-1516, 1982;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kost, G. J.
Right arrow Articles by Kenny, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kost, G. J.
Right arrow Articles by Kenny, M.

Clinical Chemistry, Vol 28, 1514-1516, Copyright © 1982 by American Association for Clinical Chemistry

Unpredictable fluctuations in transcutaneous pCO2 from capillary blood gas determinations

GJ Kost, JL Chow and M Kenny

Transcutaneous pCO2 electrode response time was optimized by use of a new electrode filling solution composed of NaCl/NaHCO3 electrolyte buffer (100 and 20 mmol/L, respectively) in an equivolume mixture of glycerol and water. The 95% response time to a step change in pCO2 was 49.9 +/- 2.8 s (mean +/- SD) when there was no spacer between the membrane and glass of the electrode. Use of this filling solution during monitoring of severely ill premature infants with cardiopulmonary disease allowed identification of large, unpredictable transient changes in transcutaneous pCO2, and therefore presumably in arterial pCO2, that occurred during capillary blood gas sampling. The changes, which occurred in 19 of 20 samplings, ranged from -1.06 + 2.53 kPa (-8 to +19 Torr). The maximum relative change observed was +29%. These results indicate that the standard protocol for capillary blood collection induces significant transient fluctuations in blood gas tensions. We believe these fluctuations decrease the reliability of capillary pCO2 values for use in clinical management in patient populations similar to ours.





HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1982 by the American Association for Clinical Chemistry.