Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 28: 1894-1898, 1982;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Giegel, J. L.
Right arrow Articles by Sheiman, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Giegel, J. L.
Right arrow Articles by Sheiman, M.

Clinical Chemistry, Vol 28, 1894-1898, Copyright © 1982 by American Association for Clinical Chemistry

Radial partition immunoassay

JL Giegel, MM Brotherton, P Cronin, M D'Aquino, S Evans, ZH Heller, WS Knight, K Krishnan and M Sheiman

In radial partition immunoassay, radial chromatography is used for performing an immunoassay. We describe the application of this technology to the measurement of digoxin in serum by enzyme immunoassay, with the entire testing procedure carried out on glass- fiber filter paper. A sample is applied to a small central area of the filter paper, where it reacts with the antibody to digoxin immobilized there. Subsequently, enzyme-labeled digoxin is applied to react with remaining antibody sites. After incubation, substrate for the enzyme is applied to the center of the reaction area and washes out any unbound label to the periphery of the paper. This step also initiates the enzyme reaction, which is quantified by front-surface fluorescence. A microprocessor-controlled automated instrument has been developed to process the filter paper matrix through the above sequence, and calculate the final result. Total testing time for digoxin is less than 7 min.


The following articles in journals at HighWire Press have cited this article:


Home page
Arch Intern MedHome page
N. B. Watts
Use of a Sensitive Thyrotropin Assay for Monitoring Treatment With Levothyroxine
Arch Intern Med, February 1, 1989; 149(2): 309 - 312.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1982 by the American Association for Clinical Chemistry.