Clinical Chemistry, Vol 29, 1937-1940, Copyright © 1983 by American Association for Clinical Chemistry
V Bhupal and RS Mani
In this simple and low-cost radioimmunoassay (RIA) procedure for thyroxin (T4), ethacridine (6,9-diamino-2-ethoxyacridine lactate; Rivanol) is used to separate antibody-bound ligand and free ligand. This acridine dye precipitates free T4 not bound to antibody. At pH 8.6, it precipitates more than 90% of 125I-labeled T4. Precipitation is rapid and is constant over the concentration range of 1 to 3 g of ethacridine per liter; time of addition and temperature are not critical. The inherent yellow color of the dye gives it the advantage of a color-coded reagent. Analysis of 65 serum samples by the proposed method and by a procedure involving polyethylene glycol gave similar results: correlation coefficient, 0.988; slope, 0.981; y-intercept, - 1.42. Within- and between-assay variations (CV) were less than 5 and 6%, respectively.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |