Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 29: 1497-1500, 1983;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chan, K. M.
Right arrow Articles by Ng, R. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chan, K. M.
Right arrow Articles by Ng, R. H.

Clinical Chemistry, Vol 29, 1497-1500, Copyright © 1983 by American Association for Clinical Chemistry

Interference by hemolysis with various methods for total calcium and its correction by trichloroacetic acid precipitation

KM Chan, C Arriaga, M Landt, CH Smith and RH Ng

We evaluated the influence of hemolysis on total serum calcium as determined with the Du Pont aca, SMA 12/60, Ektachem 400, Corning 940 EGTA titrator, and the Beckman Astra 8, comparing results with those obtained by atomic absorption spectroscopy. We find that hemoglobin does not influence calcium measurements with the SMA 12/60, Ektachem 400, or Beckman Astra 8. The presence of hemoglobin exceeding 2 to 3 g/L caused falsely high results with the aca and falsely low results with the Corning 940 titrator. Similar interferences may be observed with other titrating or colorimetric procedures that involve direct reaction of the sample with o-cresolphthalein complexone or calcein. Upon removal of the hemoglobin by precipitation with trichloroacetic acid, the values obtained with the aca and the Corning 940 EGTA titrator were similar to those measured by atomic absorption. With nonhemolyzed serum samples the acid treatment had little or no effect on the aca procedure but resulted in a positive bias of approximately 10% with the EGTA procedure. Thus we recommend this trichloroacetic acid procedure for measuring calcium in hemolyzed samples with the aca and, with certain reservations, with the EGTA titrator.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
Z. Ma, T. G. Monk, L. T. Goodnough, A. McClellan, M. Gawryl, T. Clark, P. Moreira, P. E. Keipert, and M. G. Scott
Effect of hemoglobin- and Perflubron-based oxygen carriers on common clinical laboratory tests
Clin. Chem., September 1, 1997; 43(9): 1732 - 1737.