NB Smith
In this method for detection and quantification of volatile alcohols by capillary gas chromatography, the serum sample is deproteinized, then directly injected into the gas chromatograph with 1-propanol as the internal standard. The capillary column is a 30-m bonded methylsilicone- coated, fused-silica column. With helium as the carrier gas, the injector inlet is set at a split ratio of 1/30 and the average linear velocity in the column is 25 cm/s. Injector and flame-ionization detector temperatures are 280 degrees C, oven temperature 35 degrees C. Chromatography time is less than 3 min.
The following articles in journals at HighWire Press have cited this article:
![]() |
I.-M. Jonsson, M. Verdrengh, M. Brisslert, S. Lindblad, M. Bokarewa, U. Islander, H. Carlsten, C. Ohlsson, K. S. Nandakumar, R. Holmdahl, et al. From the Cover: Ethanol prevents development of destructive arthritis PNAS, January 2, 2007; 104(1): 258 - 263. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A. Barton, P. J. Deisinger, J. C. English, J. M. Gearhart, W. D. Faber, T. R. Tyler, M. I. Banton, J. Teeguarden, and M. E. Andersen Family Approach for Estimating Reference Concentrations/Doses for Series of Related Organic Chemicals Toxicol. Sci., March 1, 2000; 54(1): 251 - 261. [Abstract] [Full Text] [PDF] |
||||
![]() |