Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 30: 1857-1860, 1984;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mullins, R. E.
Right arrow Articles by Bennett, B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mullins, R. E.
Right arrow Articles by Bennett, B.

Clinical Chemistry, Vol 30, 1857-1860, Copyright © 1984 by American Association for Clinical Chemistry

Standardized automated assay for functional alpha 1-antitrypsin

RE Mullins, RL Miller, RL Hunter and B Bennett

We describe an enzymic assay for the functional activity of alpha 1- antitrypsin, in which tosyl-Gly-Pro-Lys-p-nitroanilide acetate (Chromozym PL) is the trypsin substrate. For the indicating reaction, the concentration of substrate is about double the Km of its reaction with trypsin. This assay is an improvement over reactions involving N- alpha-benzoyl-DL-arginine p-nitroanilide as substrate, the solubility of which in water is less than the Km of its reaction with trypsin. The assay is standardized in terms of active sites of trypsin inhibited per liter of serum, with p-nitrophenyl-p'-guanidinobenzoate as the active site titrant. The results agreed well with those by an immunonephelometric assay for alpha 1-antitrypsin (r = 0.953). The within-run and run-to-run precision was approximately 3%. The results of the assay varied linearly with alpha 1-antitrypsin concentration from 0 to 60 mumol/L of serum. The minimum sample size was 50 microL, and analysis of 24 samples took less than 20 min.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
D. Roche, A. Mesner, M. Al Nakib, F. Leonard, and P. Beaune
Automated Determination of Serum {alpha}1-Antitrypsin by Antitryptic Activity Measurement
Clin. Chem., March 1, 2009; 55(3): 513 - 518.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1984 by the American Association for Clinical Chemistry.