Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 30: 1163-1167, 1984;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kemp, G. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kemp, G. J.

Clinical Chemistry, Vol 30, 1163-1167, Copyright © 1984 by American Association for Clinical Chemistry

Theoretical aspects of one-point calibration: causes and effects of some potential errors, and their dependence on concentration

GJ Kemp

The potential errors of the one-point calibration technique are described mathematically. I examine the theoretical causes and effects of analytical errors in absorbance and of errors in determining calibration concentration, and discuss techniques for minimizing their impact. The dependence of these errors on the calibration concentration and on the size of the result is considered, and some conclusions are drawn about the choice of calibration concentration.