Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 30: 1386-1388, 1984;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kroll, M. H.
Right arrow Articles by Elin, R. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kroll, M. H.
Right arrow Articles by Elin, R. J.

Clinical Chemistry, Vol 30, 1386-1388, Copyright © 1984 by American Association for Clinical Chemistry

Effect of dialysis on interference by cefoxitin with determination of creatinine

MH Kroll, C Hagengruber and RJ Elin

Interference by cefoxitin with determination of creatinine is less with the Technicon SMAC than with other commercial analytical systems. The SMAC assay involves a single-point kinetic method with dialysis, whereas most other commercial methods are multipoint kinetic without dialysis. The apparent creatinine concentration measured for aqueous solutions of cefoxitin was 73 mmol of creatinine per mole of cefoxitin with the SMAC, 135 mmol/mol with a manual method. Furthermore, we determined for the SMAC that the average fraction of creatinine dialyzed was 0.128 and for cefoxitin, 0.064. Thus, the concentration of and interference by cefoxitin in the reaction mixture for SMAC are reduced by half (i.e., the apparent creatinine concentration for cefoxitin with the manual system multiplied by 0.5 is essentially that noted with the SMAC: 68 vs 73). Thus we conclude that the diminished interference be cefoxitin with determination of creatinine by SMAC is primarily ascribable to the dialysis step.