Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 30: 1499-1501, 1984;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hoffman, K. L.
Right arrow Articles by Miles, L. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hoffman, K. L.
Right arrow Articles by Miles, L. E.

Clinical Chemistry, Vol 30, 1499-1501, Copyright © 1984 by American Association for Clinical Chemistry

Elimination of "hook-effect" in two-site immunoradiometric assays by kinetic rate analysis

KL Hoffman, GH Parsons, LJ Allerdt, JM Brooks and LE Miles

Two-site or sandwich immunoradiometric assays (IRMAs) offer theoretical advantages over competitive immunoassay systems for sensitivity, precision, and rapid incubation. The practical realization of these advantages has been limited by the phenomenon of the "high-dose hook effect," such that high concentrations of an analyte give similar responses to those of much lower concentrations. We have developed a kinetic rate monitoring IRMA system for use with the Kineti-Count 48TM, an automated kinetic radioassay analyzer, which eliminates "hook- effect" interference, thereby permitting optimal assay design for increasing sensitivity and reducing incubation time. Practical illustration of these concepts is demonstrated by a 10-min, automated, quantitative assay we developed for human choriogonadotropin. The assay can detect as little as 1.2 int. units/L and kinetically screens for the hook effect. Kinetic rate analysis of the two-site IRMA and potentially of nonisotopic counterparts permits improvements in the speed and reliability of these immunoassays.


The following articles in journals at HighWire Press have cited this article:


Home page
J Biomol ScreenHome page
P. Newton, P. Harrison, and S. Clulow
A Novel Method for Determination of the Affinity of Protein: Protein Interactions in Homogeneous Assays
J Biomol Screen, August 1, 2008; 13(7): 674 - 682.