Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 31: 92-94, 1985;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Poon, R.
Right arrow Articles by Hinberg, I. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Poon, R.
Right arrow Articles by Hinberg, I. H.

Clinical Chemistry, Vol 31, 92-94, Copyright © 1985 by American Association for Clinical Chemistry

Indican interference with six commercial procedures for measuring total bilirubin

R Poon and IH Hinberg

We have studied the effect of indican on six commercial procedures for the measurement of total bilirubin in serum. Total bilirubin measured by the Bilirubin A-Gent (Abbott) 2,4-dichlorophenyl diazonium procedure increased by 50 mg/L for each 1 mmol/L of added indican. Similarly, total bilirubin measured by the Bilirubin C-System (Boehringer Mannheim) 2,5-dichlorophenyl diazonium procedure increased by 33 mg/L per mmol/L of indican. Indican also interfered with the Micro Bilirubin Reagent Set (Harleco) Malloy-Evelyn procedure, but to a much lesser extent. The Jendrassik Bilirubin Reagent System (American Monitor) and a modified Jendrassik-Grof procedure (Hoffmann-LaRoche) adapted to the Cobas Bio analyzer were unaffected by the presence of indican. The amount of interference with the 2,5-dichlorophenyl diazonium procedure increased significantly with color development time and was twice the initial amount after 30 min. Concentrations of indican as high as 0.38 mmol/L have been found in sera of patients with renal failure, which would increase total bilirubin values measured by the first two procedures above by 19 and 12 mg/L, respectively. Users of these procedures should therefore be suspicious of unexpectedly high bilirubin values obtained with sera from patients with chronic renal disease.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
L. M. Henrich, D. E. Bruns, D. M. Haverstick, V. G. Reynolds, and J. C. Boyd
Indican Interference in Bilirubin Assays: A Classical Solution Still Applies
Clin. Chem., February 1, 2007; 53(2): 359 - 361.