Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 31: 1896-1899, 1985;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Winter, S. D.
Right arrow Articles by Lepoff, R. B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Winter, S. D.
Right arrow Articles by Lepoff, R. B.

Clinical Chemistry, Vol 31, 1896-1899, Copyright © 1985 by American Association for Clinical Chemistry

"Carryback": effect of viscous liquid controls on the preceding sample analyzed with the SMA II continuous-flow analyzer

SD Winter, C Kerr, M McAvoy, MD Gallomore, C Smith and RB Lepoff

We systematically studied the "carryback" effect of ethylene glycol- based controls on the preceding sample on an SMA II continuous-flow analyzer. Including Beckman Level 1 unassayed liquid control material as a sample lowered the 12 analyte values of the preceding sample by an average of 2.7%, Level 2 (the most viscous) by an average of 4.4%, and Level 3 by 3.2%. Water-reconstituted lyophilized control material caused no carryback effect, but lyophilized control reconstituted with 330 mL/L ethylene glycol decreased the preceding sample's results by 4.1% (average carryback). We believe that carryback is caused by the drag placed on the sample line by a viscous sample, which decreases the volume of the preceding sample that is delivered to the reagent or pre- dilution mixing coils. Our findings were confirmed on another SMA II. Limited study of a SMAC analyzer gave inconclusive results, but further evaluation of continuous-flow systems for carryback is warranted. Carryback substantially increases total analytical variability.