Clinical Chemistry, Vol 31, 2018-2019, Copyright © 1985 by American Association for Clinical Chemistry
DB Barnes, GF Pierce, D Lichti, M Landt, J Koenig and KM Chan
We evaluated the effect of dextran on values for total protein in serum as measured by the biuret method with five widely used automated instruments: the American Monitor Parallel; the Du Pont aca II; the Roche Cobas-Bio; the Kodak Ektachem 400; and the Beckman Astra 8. Dextran concentrations as great as 25 or 30 g/L had relatively little or no influence on total protein measurements by the latter three instruments. Dextran concentrations exceeding 6 g/L caused falsely low results with the aca, whereas the Parallel gave falsely high results when the dextran concentration exceeded 2 g/L. The aca total protein procedure could be protected from the interference by dextran concentrations up to 30 g/L by injecting 0.4-0.8 mL of ethylene glycol directly into the reagent pack before sampling. However, we could not eliminate the interference with the Parallel procedure by any simple means; we thus recommend that it not be used for measuring total protein in serum samples from patients who are being treated with dextran.