Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 31: 196-201, 1985;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lensmeyer, G. L.
Right arrow Articles by Fields, B. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lensmeyer, G. L.
Right arrow Articles by Fields, B. L.

Clinical Chemistry, Vol 31, 196-201, Copyright © 1985 by American Association for Clinical Chemistry

Improved liquid-chromatographic determination of cyclosporine, with concomitant detection of a cell-bound metabolite

GL Lensmeyer and BL Fields

This unique extraction and isocratic "high-performance" liquid chromatographic method for measuring cyclosporine (CsA) in blood involves a Zorbax cyanopropyl analytical column maintained at 58 degrees C, with detection at 214 nm, and recycling of the water:acetonitrile mobile phase for improved long-term column stability and efficiency. Routinely, 1.0 mL of serum, plasma, or whole blood is diluted with water:acetonitrile (70:30) and applied to a disposable solid-phase cyanopropyl column to rapidly extract the drug and the internal standard cyclosporin D (CsD). Analytical recovery for this step averages 90% with whole blood and 98% with serum and plasma. Between-run CVs were 6.5 and 2.6% for means of 104 and 1128 micrograms/L, respectively. The standard curve is linear up to 1600 micrograms/L. The minimum detection limit is 10 to 15 micrograms/L. No interferences from endogenous substances or other drugs were found. In addition, a compound cross reacting with the Sandoz radioimmunoassay antibody was isolated from patients' samples with the present procedure and was tentatively identified as a CsA metabolite(s). It appears to be highly partitioned on blood cells, very little being detected in the serum or plasma. In a comparison with RIA, correlation coefficients were 0.828 and 0.652 for serum and whole blood, respectively. Results from a 12-h pharmacokinetic study in which different sample types were analyzed by RIA and liquid chromatography further exemplified major discrepancies between types of CsA determinations.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
R. Rej
Clinical Chemistry through Clinical Chemistry: A Journal Timeline
Clin. Chem., December 1, 2004; 50(12): 2415 - 2458.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1985 by the American Association for Clinical Chemistry.