Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 31: 1229-1231, 1985;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by McPhaul, L.
Right arrow Articles by Eckfeldt, J. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by McPhaul, L.
Right arrow Articles by Eckfeldt, J. H.

Clinical Chemistry, Vol 31, 1229-1231, Copyright © 1985 by American Association for Clinical Chemistry

A 2,4-dichlorophenyl diazonium-based method for total bilirubin without interference from indican in uremic sera

L McPhaul, M Kershaw, D Tilque and JH Eckfeldt

Comparison of total bilirubin quantification by a 2,4-dichlorophenyl diazonium method (2,4-DCPD) with a Jendrassik-Grof type of method showed excellent correlation for randomly selected sera. However, sera from uremic patients on chronic hemodialysis showed a marked positive bias for the 2,4-DCPD result as compared with the Jendrassik-Grof result. The mean difference was 5.3 mg/L, and resulted in about 20% of the hemodialysis patients having bilirubin values greater than 13 mg/L, the upper limit of our reference range. Indican in uremic sera reportedly reacts with certain diazo reagents, so we investigated indican's reactivity in the above methods. In vitro addition of indican caused no interference in the Jendrassik-Grof method, but produced a significant positive interference in the 2,4-DCPD method, 1 mmol of indican per liter appearing as about 36 mg of total bilirubin per liter. Long reaction times with the 2,4-DCPD reagent accentuate the problem. By shortening the reaction time with the 2,4-DCPD reagent to 1.7 min, we find that the indican interference can be eliminated, without affecting quantification of total bilirubin in either normal or uremic sera.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
L. M. Henrich, D. E. Bruns, D. M. Haverstick, V. G. Reynolds, and J. C. Boyd
Indican Interference in Bilirubin Assays: A Classical Solution Still Applies
Clin. Chem., February 1, 2007; 53(2): 359 - 361.