Clinical Chemistry, Vol 32, 1873-1878, Copyright © 1986 by American Association for Clinical Chemistry
WD Odell, J Griffin and R Zahradnik
We have developed a sensitive, specific, noncompetitive, sandwich-type radioimmunoassay for human thyrotropin (hTSH), which can be performed in 30 min. The assay involves two monoclonal antibodies, selected for high affinity and specificity and also for reaction against antigenic sites on hTSH that are distal from each other. One of these antibodies is labeled with 125I; the other is conjugated covalently to biotin. Polystyrene beads were also conjugated covalently to biotin. After conjugation, the beads were incubated with avidin. These beads represent a rapid, simple method for separating hTSH-bound antibody from free antibody. The biotin-antibody-hTSH-125I-labeled antibody complexes bind to the beads and hTSH concentration is directly related to counts per minute. This assay can detect hTSH at a concentration of 0.06 milli-unit/L in serum.
The following articles in journals at HighWire Press have cited this article:
![]() |
J. Deinum, F. H.M. Derkx, and M. A.D.H. Schalekamp Improved Immunoradiometric Assay for Plasma Renin Clin. Chem., June 1, 1999; 45(6): 847 - 854. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. H. J. Danser, C. A. M. van Kesteren, W. A. Bax, M. Tavenier, F. H. M. Derkx, P. R. Saxena, and M. A. D. H. Schalekamp Prorenin, Renin, Angiotensinogen, and Angiotensin-Converting Enzyme in Normal and Failing Human Hearts : Evidence for Renin Binding Circulation, July 1, 1997; 96(1): 220 - 226. [Abstract] [Full Text] |
||||