Clinical Chemistry, Vol 32, 1942-1943, Copyright © 1986 by American Association for Clinical Chemistry
GC Zucchelli, A Pilo, MR Chiesa, S Masini and C Baccini
Between-laboratory agreement of CEA determinations collected in a multicenter study has been substantially improved (CV decreased from 60.4 to 21.6%) by converting the results from mass concentration units (ng/mL) of local kit standards to int. units/L of the international standard first IRP 73/601. Conversion factors for the kits most used were computed, during the same survey, from results of two analytical- recovery experiments in which control samples, supplemented with the international standard, were analyzed by participating laboratories (n = 96).
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |