Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 32: 768-772, 1986;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Pankey, S.
Right arrow Articles by Singh, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pankey, S.
Right arrow Articles by Singh, P.

Clinical Chemistry, Vol 32, 768-772, Copyright © 1986 by American Association for Clinical Chemistry

Quantitative homogeneous enzyme immunoassays for amitriptyline, nortriptyline, imipramine, and desipramine

S Pankey, C Collins, A Jaklitsch, A Izutsu, M Hu, M Pirio and P Singh

We describe specific EMIT homogeneous enzyme immunoassays for amitriptyline, nortriptyline, imipramine, and desipramine in patients' serum samples. Before analysis, an easily performed extraction step involving the use of 500 microL of sample and a 1-mL disposable column eliminates cross-reacting polar metabolites. The range of the standard curve for the first three drugs is 25 to 250 micrograms/L, and for desipramine is 50 to 500 micrograms/L. Within-run and between-run CVs are less than 10% throughout the range of the assays. Results for patients' samples obtained by this method and by "high-performance" liquid chromatography compare well, showing a slope range of 0.94-1.04 and correlation coefficients ranging from 0.93 to 0.96, depending on the assay.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
M. W. Linder and P. E. Keck Jr.
Standards of laboratory practice: antidepressant drug monitoring
Clin. Chem., May 1, 1998; 44(5): 1073 - 1084.