Clinical Chemistry, Vol 32, 768-772, Copyright © 1986 by American Association for Clinical Chemistry
S Pankey, C Collins, A Jaklitsch, A Izutsu, M Hu, M Pirio and P Singh
We describe specific EMIT homogeneous enzyme immunoassays for amitriptyline, nortriptyline, imipramine, and desipramine in patients' serum samples. Before analysis, an easily performed extraction step involving the use of 500 microL of sample and a 1-mL disposable column eliminates cross-reacting polar metabolites. The range of the standard curve for the first three drugs is 25 to 250 micrograms/L, and for desipramine is 50 to 500 micrograms/L. Within-run and between-run CVs are less than 10% throughout the range of the assays. Results for patients' samples obtained by this method and by "high-performance" liquid chromatography compare well, showing a slope range of 0.94-1.04 and correlation coefficients ranging from 0.93 to 0.96, depending on the assay.
The following articles in journals at HighWire Press have cited this article:
![]() |
M. W. Linder and P. E. Keck Jr. Standards of laboratory practice: antidepressant drug monitoring Clin. Chem., May 1, 1998; 44(5): 1073 - 1084. [Abstract] [Full Text] [PDF] |
||||