Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 32: 850-853, 1986;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Warwick, W. J.
Right arrow Articles by Strominger, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Warwick, W. J.
Right arrow Articles by Strominger, D.

Clinical Chemistry, Vol 32, 850-853, Copyright © 1986 by American Association for Clinical Chemistry

Evaluation of a cystic fibrosis screening system incorporating a miniature sweat stimulator and disposable chloride sensor

WJ Warwick, NN Huang, WW Waring, AG Cherian, I Brown, E Stejskal-Lorenz, WH Yeung, G Duhon, JG Hill and D Strominger

A new sweat test (CF Indicator; Medtronic, Inc.) for cystic fibrosis (CF) features a compact, portable configuration of electrodes that dispense pilocarpine for iontophoresis. A disposable chloride sensor patch absorbs a specified volume of sweat, in which the chloride concentration is immediately determined as less than 40, 40-60, or greater than 60 mmol/L. We assessed the performance of the system in a five-center study, in relation to the clinical diagnosis and to the Gibson-Cooke sweat test (GCST) as a control test. With sweat chloride concentrations of less than or equal to 40 mmol/L defined as normal and greater than 40 mmol/L as indicating persons at risk for CF, the new system showed 91% specificity and 100% sensitivity for CF, as compared with 92.8% and 100%, respectively, for the GCST. When we used sweat chloride concentrations of less than or equal to 60 mmol/L as probably normal and greater than 60 mmol/L as probably indicative of CF, the new system showed a 99.1% specificity and 98.6% sensitivity, vs 97.8% specificity and 97.9% sensitivity for the GCST test. In both procedures, occasionally insufficient sweat was collected, and this appeared related to the age of the subject. We conclude that the new sweat test system is potentially useful in physicians' offices, in clinics, and similar settings.


The following articles in journals at HighWire Press have cited this article:


Home page
ChestHome page
M. A. Martinez-Garcia, J.-J. Soler-Cataluna, M. Perpina-Tordera, P. Roman-Sanchez, and J. Soriano
Factors Associated With Lung Function Decline in Adult Patients With Stable Non-Cystic Fibrosis Bronchiectasis
Chest, November 1, 2007; 132(5): 1565 - 1572.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1986 by the American Association for Clinical Chemistry.